hameg-hmo: When setting slope, also set trigger type to edge.
When setting the type of slope for the edge trigger function, also set the trigger type to edge because it is not necessarily configured that way and therefore such functionality might fail to work properly! This fixes parts of bug #1328.
This commit is contained in:
parent
a9f3fa0548
commit
c8ecfa6ea4
|
@ -45,7 +45,7 @@ static const char *hameg_scpi_dialect[] = {
|
|||
[SCPI_CMD_GET_DIG_POD_STATE] = ":POD%d:STAT?",
|
||||
[SCPI_CMD_SET_DIG_POD_STATE] = ":POD%d:STAT %d",
|
||||
[SCPI_CMD_GET_TRIGGER_SLOPE] = ":TRIG:A:EDGE:SLOP?",
|
||||
[SCPI_CMD_SET_TRIGGER_SLOPE] = ":TRIG:A:EDGE:SLOP %s",
|
||||
[SCPI_CMD_SET_TRIGGER_SLOPE] = ":TRIG:A:TYPE EDGE;:TRIG:A:EDGE:SLOP %s",
|
||||
[SCPI_CMD_GET_TRIGGER_SOURCE] = ":TRIG:A:SOUR?",
|
||||
[SCPI_CMD_SET_TRIGGER_SOURCE] = ":TRIG:A:SOUR %s",
|
||||
[SCPI_CMD_GET_DIG_CHAN_STATE] = ":LOG%d:STAT?",
|
||||
|
|
Loading…