Update the Hameg/Rohde&Schwarz HMO driver (hameg-hmo) so that it
is possible to configure the logic threshold for digital signals.
The user can get or set the logic threshold configuration using
the channel group POD0 (and/or POD1 where available), for example:
sigrok-cli --driver hameg-hmo --get logic_threshold -g POD0
sigrok-cli --driver hameg-hmo --config logic_threshold=TTL --set -g POD0
sigrok-cli --driver hameg-hmo --get logic_threshold_custom -g POD0
sigrok-cli --driver hameg-hmo --config logic_threshold_custom=0.7 --set -g POD0
Update the default serial port options for Rohde&Schwarz and
Hameg mixed-signal oscilloscope devices connected through USB.
Also, remove misplaced and unused serial port configuration option.
This patch complements fa3d104f17
in terms of updating the USB PIDs for new devices (HMO series).
This fixes parts of bug #1321.
If no serial port option is specified on the command-line using the
"serialcomm" driver option, but the device is connected through USB
and it requires a known default serial port option, then use it in
order to avoid data corruption or even worse problems.
Note: the easiest way to reproduce data corruption on HMO3000 series
is to start an analog data acquisition (e.g. on channel CH1) after
switching from Ethernet mode to USB VCP mode (HO732 USB/Ethernet interface).
This fixes parts of bug #1321.
Correctly set the length of the buffer used to hold the SCPI response
from the device containing the binary acquisition data.
If a timeout occurs, truncate the buffer and send the partial response
from the device instead of getting stuck on timeouts!
Thanks to Stefan Brüns for reviewing the first version of this patch
and spotting out a serious problem with it.
This fixes bug #1323.
At the moment only the maximum number of frames to be acquired can be
configured for the Hameg/Rohde&Schwarz HMO mixed-signal oscilloscope
series driver (hameg-hmo).
This patch adds support to configure the number of samples to acquire
in both analog and digital (logic) mode.
This patch introduces the support for 16 digital (logic) channels for the
following oscilloscope models: HMO3032, HMO3042, HMO3052 and HMO3522
(previously only 8 digital channels were supported, i.e. only 1 POD).
This patch takes care of removing an invalid product model (HMO2522)
and adds a missing product model (HMO3522) in the hameg-hmo driver,
thus extending the number of supported oscilloscope models.
This fixes bug #1322.
The trigger range/mask "compression" procedure is apparently not
required and breaks triggering on a low line state.
This has been verified to fix the issue on a Hantek 4032L with FPGA
version 0x4303. It is possible that the procedure mentioned above
might be required for other FPGA versions, though that is unknown.
If you experience trigger issues with other FPGA versions, please
contact us for further debugging and testing.
This fixes bug #1402.
The Brymen BM86x supports up to two temperature probes. The dual display
can show individual temperatures of the two probes or differences. The
previous implementation "detected" a value of zero degrees when no probe
was attached and the display showed dash lines. When cycling assignments
of probes to displays, some valid combinations did not result in values
shown by the libsigrok driver.
An implementation detail of the formerly separate brymen-bm86x driver
was lost during recent migration of the dmm/bm86x parser into the
serial-dmm driver. When the meter's temperature function is selected,
it's essential to inspect the primary display's flags and digits, to
determine the secondary display's quantity and unit. Previous versions
never bothered to explicitly check for the "----" digits text, but the
combination of the primary's last digit showing C/F as well as the
binary C/F flags before the value provided hints which the secondary
displays group of digits and flags did not.
This fixes bug #1394.
When the list of all connections gets created which are supported by the
HID serial transport, items contain a "hid/ch9325/raw=/dev/hidraw3" path
and a "1a86:e008" pair of vendor and product IDs.
Separate the VID/PID pair by a period not a colon, so that --list-serial
output immediately becomes usable with "--driver <name>:conn=<spec>"
invocations. Eliminate the necessity to adjust clipboard context by the
user. This improves usability in cases where not a single connection
gets addressed, but a group of connections gets specified by ambiguous
conn= specs.
$ sigrok-cli -d uni-t-ut32x:conn=1a86.e008 --scan
Some meters require the reception of a request before they provide
acquisition data. Add support for the chip driver's .packet_request()
routine, and timeout handling in the serial-lcr driver. This follows
the serial-dmm model.
Allow LCR chip drivers to specify custom printf() formats for their
channel names. Default to "P1" etc in the absence of format specs.
This implementation is similar to serial-dmm.
The EEVBlog 121GW meter support always registers the three-displays
parse routine with the serial-dmm device driver. The single-display
routine need not be public. Adjust the visibility.
Reduce indentation for a continued line in a nearby declaration
while we are here.
Bug #1390 reports that "!desc" is always true (should be: false?). But
the actual problem would be that 'desc' is _not_ NULL when none of the
supported chips' VID:PID matched (FT232H happens to "get found" then,
erroneously).
Add a sentinel to the table of supported chips, such that 'desc' becomes
NULL upon mismatch, and the error path is entered.
The serial-dmm entry for Victor DMMs is able to communicate to either
geniune COM ports (RS232 or USB CDC) or the obfuscating USB HID cables.
The victor-dmm driver became obsolete and was removed. Rename the
serial-dmm entry, remove the no longer needed "-ser" suffix. Suggest a
default connection via the obfuscating USB HID cable, accept user
provided overrides for regular serial cables (same behaviour as before,
connection spec for serial ports keeps being mandatory).
Remove the victor-dmm device driver. Its functionality is contained in
the Victor specific serial-over-HID transport, the FS9922 DMM parser,
and the serial-dmm device driver. The additional implementation became
obsolete.
Introduce a serial transport which undoes the Victor DMM cable's
obfuscation to the DMM chip's original data packet. Which allows to
re-use the existing FS9922 support code, obsoleting the victor-dmm
device driver.
This undoes the essence of commit bf700f679a, which introduced the
default conn= spec. Which improved usability: Users just select the
device and need not specify the connection. But also resulted in the
UT32x thermometer driver's probe to "succeed" as soon as the WCH CH9325
chip was found, no data check was involved in the scan. Unfortunately
this chip is also used in the popular UT-D04 cable, and thus there were
many false positives.
Remove the src/hardware/brymen-bm86x/ hierarchy of source files. Its
functionality has moved to the bm86x packet parser and the serial-dmm
device driver.
Request packets from the Brymen BM86x meter much faster. The previous
implementation in the separate driver used to immediately send another
request when a measurement arrived, with a 10ms granularity in the poll
routine, and a 500ms timeout between requests.
Considering the meter's update rate, stick with the 500ms timeout, but
increase the maximum request rate to 10 per second, with a minimum of 2
per second. This receives measurement data at the meter's capability
(compare DC and AC modes, seems to automatically adjust to the internal
operation, and match the display update rate).
Move Brymen BM86x specific packet parse logic to a new src/dmm/bm86x.c
source file, and register the routines with the serial-dmm driver's list
of supported devices. Which obsoletes the src/hardware/brymen-bm86x/
hierarchy.
This implementation differs from the previous version: The parse routine
gets called multiple times after one DMM packet was received. Values for
the displays get extracted in separate invocations, the received packet
is considered read-only. Unsupported LCD segment combinations for digits
get logged. Low battery gets logged with higher severity -- the validity
of measurement values is uncertain after all. The parse routine uses
longer identifiers. Packet reception uses whichever serial transport is
available (need no longer be strictly USB HID nor libusb based). All
features of the previous implementation are believed to still be present
in this version.
This configuration queries measurement values each 0.5 seconds and
re-sends a not responded to request after 1.5 seconds. Which follows the
combination of the vendor's suggested flow (frequency) and the previous
implementation's timeout (3x 500ms). This implementation does not try to
re-connect to the HID device for each measurement, and neither checks
for the 4.0 seconds timeout (vendor's suggested flow). Local experiments
work without these.
The src/hardware/brymen-bm86x/ source code contains specific support for
the Brymen BM86x devices, and directly depends on the libusb library.
Rename the registered device (append the "-usb" suffix) before adding
BM86x support to the serial-dmm driver.
The Brymen BU-86X infrared adapters are sold with BM869s meters. Raw
streams of data bytes get communicated by means of HID reports with
report number 0 and up to 8 data bytes each. Communication parameters
are fixed and need no configuration.
Some meters which are supported by the serial-dmm driver don't strictly
require the user's COM port specification. When a known (usually bundled,
or even builtin) cable type is used, we can provide a default conn= spec
and thus improve usability. Prepare the DMM_CONN() macro, accept user
overrides.
The 'conn' field in the device context and the CONN values in the
declarations of supported DMM models seemed inappropriate. They specify
the communication parameters (UART frame format and bitrate), not the
connection (port name). Adjust the respective identifiers.
Also rephrase the evaluation logic. Instead of checking for the absence
of user specs and optionally assigning a fallback value, just preset
from defaults and override from user specs when present. This simplifies
the logic (eliminates a check).