Prefer the common conversion helper for little endian 16bit signed data.
The previous local implementation only worked for positive values, and
yielded incorrect results for negative temperatures.
This fixes bug #1463.
Since we've now seen lots of devices in the wild that come with the
"HCS-" prefix in the ID, it's probably safe to assume all of them
could have it.
This fixes bug #1530.
The previous implementation got stuck in an infinite loop when data
acquisition started, but the device got disconnected before the data
acquisition terminates. An implementation detail ignored communication
errors, and never saw the expected condition that was required to
continue in the sample download sequence. Unbreak that code path.
Even though the devices/websites/manuals usually say 0..30V, the
hardware actually accepts up to 31V, both via serial as well as by
simply rotating the knob on the device (and our driver already
reflects that).
The same is true for current, it's usually 0..5A as per docs, but many
(probably all) devices accept 5.1A via serial and knob.
Thus, set the max current of all devices to 5.1A (or 3.1A for 3A
devices). We're assuming they all have this property, and we've seen
this in practice on at least three different versions of the device.
On a recently acquired Korad KA3005P power supply, the ID supplied by the
device is not known by libsigrok.
$ sigrok-cli --driver=korad-kaxxxxp:conn=/dev/ttyACM0 --scan
sr: korad-kaxxxxp: Unknown model ID 'KORAD KA3005P V4.2' detected, aborting.
This fixes bug #1522.
Thanks to bitaround@gmail.com for the amperage fix.
The DMMs report as an event to which mode the user switched (by turning the
rotary switch): "*0", "*1", etc.
Most other DMMs have few modes, but the U127x DMMs have up to 11 different
modes (i.e., "*10" is a valid event).
Add a comment on the logic which skips the upper 64 bytes of a 512 bytes
chunk in the Asix Sigma's sample memory. Move the initial assignment and
the subsequent update from a value which was retrieved from a hardware
register closer together for awareness during maintenance. Pre-setting a
high position value that will never match when the feature is not in use
is very appropriate.
Adjust the sigma_read_pos() routine to handle triggerpos identically to
stoppos. The test condition's intention is to check whether a decrement
of the position ends up in the meta data section of a chunk. The previous
implementation tested whether a pointer to the position variable ended in
0x1ff when decremented -- which is unrelated to the driver's operation.
It's assumed that no harm was done because the trigger feature is
unsupported (see bug #359).
This silences the compiler warning reported in bug #1411.
The handler for fluke 18x and 28x DMMs allocates several data
structures on the stack that are used after they have been freed when
creating a data feed packet.
Restructure the code so that all handlers send their own packets. As a
bonus, this avoid a couple of small heap allocations.
By separating the variables that holds the get and set values, the output
state, OVP and OCP can now be set while the acquisition is running.
Also some variables are named more clearly.
The communication with the rdtech power supplies is not very reliable,
especially during the start. Because of that, the driver tries to read the
modbus registers up to three times.
Nevertheless there is the chance, that the communication fails.
Introduce quirks support for devices which provide incomplete metadata.
Add conservative logic to unbreak the Logic Shrimp. Amend previously
received information when it was incomplete, but don't interfere if a
future firmware version fixes the issue.
Without this change, the device gets detected but "has zero channels"
and would be unusable. Because when a device provides metadata, these
details are used exclusively, no fallbacks apply.
The Mooshimeter driver uses mem*() and str*() library calls. Include the
<string.h> header file to silence compiler warnings.
...
./src/hardware/mooshimeter-dmm/protocol.c: In function 'lookup_tree_path':
./src/hardware/mooshimeter-dmm/protocol.c:275:3: warning: implicit declaration of function 'strchr' [-Wimplicit-function-declaration]
end = strchr(path, ':');
^
./src/hardware/mooshimeter-dmm/protocol.c:275:9: warning: incompatible implicit declaration of built-in function 'strchr' [enabled by default]
end = strchr(path, ':');
^
...
This adds support for the Mooshim Engineering BLE based Mooshimeter.
Because the meter requires raw BLE packets, the driver uses the BLE
layer directly. Since the meter has no physical way of configuring it,
the actual configuration is set entirely with sigrok device options.
This adds basic support for the Rigol MSO5000 series. It has
the same problems as the DS4000 series: Live capture provides
one digital channel per byte. Buffered memory returns the data
compressed (one byte has 8 digital channels), but the banks are
read separately. It's not possible to read uint16.
Introduce the lcr/vc4080.c source file which implements the parser for
the serial data stream of the Voltcraft 4080 LCR meter. Add the meter to
the list of supported devices in the serial-lcr driver, as well as the
PeakTech 2165 LCR meter which is another compatible device.
This implementation contains a workaround for USB based serial cables
which seem to suffer from incomplete parity handling (observed with the
FT232R based PeakTech cable). Similar approaches were seen in existing
DMM drivers.
This implementation supports the main and secondary displays. The D and Q
"displays" which are communicated in the serial packets appear unreliable
and redundant, users can have the D and Q values shown in the supported
displays.
Commit cb5cd1538f introduced packet request support in the serial-lcr
device driver. Calls were added to the detection of the device's
presence, and the periodic acquisition of measurement data. Add another
call to the device configuration retrieval that follows the presence
detection, without it communication timed out with no data received.
Also slightly raise the timeout for this device configuration gathering
phase. With one second sharp, the VC4080 was detected, but getting its
current configuration kept failing. This device's serial communication
is extra slow (1200 bps) and the packets are rather large (39 bytes).
Which made the stream detect's receive routine stop checking for the
availability of more data while a packet was being received.
Move the initial device probe (LCR packet validity check), the creation
of the device instance after successful probe, and the subsequent packet
inspection after resource allocation out of the scan routine. This shall
improve readability of the serial-lcr driver's probe logic, and reduces
diffs when handling of multiple connections gets added later.
Add a developer comment, the serial-lcr driver needs to handle multiple
connections when the conn= spec is ambiguous (multiple cables of the
same type, with the same VID:PID).
According to the latest available version of the manual, as
for the RTB2000 and RTM3000 series, the RTA4000 series also
uses a slightly different dialect than other previously
supported models, in particular when it comes to the POD
(logic channel groups) handling.
I do not have such model available for testing therefore, as
for the RTB2000 and RTM3000 support recently introduced, I do
not know whether or not the RTA4000 also understands the
existing dialect. In doubt, the new official dialect is
implemented by this patch.
According to the latest available version of the manual, they
both use a slightly different dialect than currently supported
models, in particular when it comes to the POD (logic channel
groups) handling.
I do not have any of the above models available for testing
therefore I do not know whether or not they also understand
the existing dialect. In doubt, the new official dialect is
implemented by this patch.
Update the oscilloscope state with new settings only after
they have been successfully stored in the device to avoid
an inconsistent state in case of SCPI SET command failure.
The hameg-hmo driver returns an incorrect sample rate: to reproduce this
bug, set the maximum sample rate from the ACQUIRE menu.
This patch fixes the driver so that the correct sample rate is returned.
During the initial configuration phase of the hameg-hmo driver
only send an OPC command if a SCPI command has been previously
sent to the device so that bogus SCPI timeouts are avoided.
The current starting index for the POD name is currently wrong as it is zero.
The official POD numbering starts instead at 1 (see device panel, buttons
and manual), so the current index used for message printing and groups
naming in the driver needs to be incremented by one.
Avoid double memory freeing leading to segmentation fault in when a SCPI
command fails to get a string due conditions such as a timeout or an invalid
command.
When setting the type of slope for the edge trigger function, also set the
trigger type to edge because it is not necessarily configured that way and
therefore such functionality might fail to work properly!
This fixes parts of bug #1328.
Update the Hameg/Rohde&Schwarz HMO driver (hameg-hmo) so that it
is possible to configure the logic threshold for digital signals.
The user can get or set the logic threshold configuration using
the channel group POD0 (and/or POD1 where available), for example:
sigrok-cli --driver hameg-hmo --get logic_threshold -g POD0
sigrok-cli --driver hameg-hmo --config logic_threshold=TTL --set -g POD0
sigrok-cli --driver hameg-hmo --get logic_threshold_custom -g POD0
sigrok-cli --driver hameg-hmo --config logic_threshold_custom=0.7 --set -g POD0
Update the default serial port options for Rohde&Schwarz and
Hameg mixed-signal oscilloscope devices connected through USB.
Also, remove misplaced and unused serial port configuration option.
This patch complements fa3d104f17
in terms of updating the USB PIDs for new devices (HMO series).
This fixes parts of bug #1321.
At the moment only the maximum number of frames to be acquired can be
configured for the Hameg/Rohde&Schwarz HMO mixed-signal oscilloscope
series driver (hameg-hmo).
This patch adds support to configure the number of samples to acquire
in both analog and digital (logic) mode.
This patch introduces the support for 16 digital (logic) channels for the
following oscilloscope models: HMO3032, HMO3042, HMO3052 and HMO3522
(previously only 8 digital channels were supported, i.e. only 1 POD).
This patch takes care of removing an invalid product model (HMO2522)
and adds a missing product model (HMO3522) in the hameg-hmo driver,
thus extending the number of supported oscilloscope models.
This fixes bug #1322.
The trigger range/mask "compression" procedure is apparently not
required and breaks triggering on a low line state.
This has been verified to fix the issue on a Hantek 4032L with FPGA
version 0x4303. It is possible that the procedure mentioned above
might be required for other FPGA versions, though that is unknown.
If you experience trigger issues with other FPGA versions, please
contact us for further debugging and testing.
This fixes bug #1402.
Some meters require the reception of a request before they provide
acquisition data. Add support for the chip driver's .packet_request()
routine, and timeout handling in the serial-lcr driver. This follows
the serial-dmm model.
Allow LCR chip drivers to specify custom printf() formats for their
channel names. Default to "P1" etc in the absence of format specs.
This implementation is similar to serial-dmm.
Bug #1390 reports that "!desc" is always true (should be: false?). But
the actual problem would be that 'desc' is _not_ NULL when none of the
supported chips' VID:PID matched (FT232H happens to "get found" then,
erroneously).
Add a sentinel to the table of supported chips, such that 'desc' becomes
NULL upon mismatch, and the error path is entered.
The serial-dmm entry for Victor DMMs is able to communicate to either
geniune COM ports (RS232 or USB CDC) or the obfuscating USB HID cables.
The victor-dmm driver became obsolete and was removed. Rename the
serial-dmm entry, remove the no longer needed "-ser" suffix. Suggest a
default connection via the obfuscating USB HID cable, accept user
provided overrides for regular serial cables (same behaviour as before,
connection spec for serial ports keeps being mandatory).
Remove the victor-dmm device driver. Its functionality is contained in
the Victor specific serial-over-HID transport, the FS9922 DMM parser,
and the serial-dmm device driver. The additional implementation became
obsolete.
This undoes the essence of commit bf700f679a, which introduced the
default conn= spec. Which improved usability: Users just select the
device and need not specify the connection. But also resulted in the
UT32x thermometer driver's probe to "succeed" as soon as the WCH CH9325
chip was found, no data check was involved in the scan. Unfortunately
this chip is also used in the popular UT-D04 cable, and thus there were
many false positives.
Remove the src/hardware/brymen-bm86x/ hierarchy of source files. Its
functionality has moved to the bm86x packet parser and the serial-dmm
device driver.
Request packets from the Brymen BM86x meter much faster. The previous
implementation in the separate driver used to immediately send another
request when a measurement arrived, with a 10ms granularity in the poll
routine, and a 500ms timeout between requests.
Considering the meter's update rate, stick with the 500ms timeout, but
increase the maximum request rate to 10 per second, with a minimum of 2
per second. This receives measurement data at the meter's capability
(compare DC and AC modes, seems to automatically adjust to the internal
operation, and match the display update rate).
Move Brymen BM86x specific packet parse logic to a new src/dmm/bm86x.c
source file, and register the routines with the serial-dmm driver's list
of supported devices. Which obsoletes the src/hardware/brymen-bm86x/
hierarchy.
This implementation differs from the previous version: The parse routine
gets called multiple times after one DMM packet was received. Values for
the displays get extracted in separate invocations, the received packet
is considered read-only. Unsupported LCD segment combinations for digits
get logged. Low battery gets logged with higher severity -- the validity
of measurement values is uncertain after all. The parse routine uses
longer identifiers. Packet reception uses whichever serial transport is
available (need no longer be strictly USB HID nor libusb based). All
features of the previous implementation are believed to still be present
in this version.
This configuration queries measurement values each 0.5 seconds and
re-sends a not responded to request after 1.5 seconds. Which follows the
combination of the vendor's suggested flow (frequency) and the previous
implementation's timeout (3x 500ms). This implementation does not try to
re-connect to the HID device for each measurement, and neither checks
for the 4.0 seconds timeout (vendor's suggested flow). Local experiments
work without these.
The src/hardware/brymen-bm86x/ source code contains specific support for
the Brymen BM86x devices, and directly depends on the libusb library.
Rename the registered device (append the "-usb" suffix) before adding
BM86x support to the serial-dmm driver.
Some meters which are supported by the serial-dmm driver don't strictly
require the user's COM port specification. When a known (usually bundled,
or even builtin) cable type is used, we can provide a default conn= spec
and thus improve usability. Prepare the DMM_CONN() macro, accept user
overrides.
The 'conn' field in the device context and the CONN values in the
declarations of supported DMM models seemed inappropriate. They specify
the communication parameters (UART frame format and bitrate), not the
connection (port name). Adjust the respective identifiers.
Also rephrase the evaluation logic. Instead of checking for the absence
of user specs and optionally assigning a fallback value, just preset
from defaults and override from user specs when present. This simplifies
the logic (eliminates a check).
Keep the ES51919 chip support in the src/lcr/ directory, and move device
driver specific code to the src/hardware/serial-lcr/ directory. Implement
the same driver layout for LCR meters as is used for DMM devices.
This also addresses a few issues in the serial-lcr driver: Unbreak --get
and --show, do process a few LCR packets after probing the device, to
gather current parameter values. Keep sending meta packets when these
parameters change during acquisition, like the previous implementation
did. Use common code for frame/time limits.
Note that although LCR meters usually operate with AC to classify L/C/R
components, one of the officially supported modes is DC resistance.
Which means that an output frequency of 0 is not just a fallback when
packet parsing fails, it's also a regular value of the freq parameter.
List all supported frequencies including DC in strict numerical order.
Although all currently supported devices use the same ES51919 chip, the
implementation is prepared to support other devices which use different
LCR meter chips as well. The list of known equivalent circuit models and
output frequencies is kept in src/lcr/ chip support. It's assumed that
one LCR packet communicates the data for all channels/displays similar
to the serial-dmm driver implementation.
The serial communication timing parameters during probe get determined
from earlier serial port configuration, which obsoletes the redundant
'baudrate' parameter, and eliminates potential inconsistency between
user specified parameters and builtin default values.
The serial communication timing parameters during probe get determined
from earlier serial port configuration, which obsoletes the redundant
'baudrate' parameter, and eliminates potential inconsistency between
user specified parameters and builtin default values.
The serial_stream_detect() routine needs to estimate the time which is
needed to communicate a given amount of data. Since the serial port got
opened and configured before, the serial communication parameters are
known, and callers need not redundantly specify the bit rate.
Register another driver for the UNI-T UT612 LCR meter, which is based on the
ES51919/ES51920 chipset, too. This device had been usable before when the
internal UART connection was made accessible (read: with a hack). It became
officially supported in unmodified form with the addition of transparent
serial over HID support for SiLabs CP2110 chips.
Switch the UT32x driver from running specific USB transfers to generic
serial communication. Preset the bitrate and frame format, but allow for
user specified overrides as well. Default to the WCH CH9325 HID chip,
but allow for overrides or more specific selection so that users can
resolve ambiguities with multiple cables.
The switch from libusb to hidapi removes a limitation that specifically
was reported for the Mac platform. The serial-over-HID variant should
now work as well. See bug #555.
Drop the background transfers. Stick with a local acquisition stop
routine, because a STOP request needs to get sent to the device. Reduce
the receive buffer size such that either blocking or non-blocking calls
will work. The additional flexibility of the buffer handling and packet
processing does not harm.
Eliminate a direct libserialport dependency in the OLS device driver.
Use libsigrok's internal serial layer's API instead to check for the
availability of receive data.
The rohde-schwarz-sme-0x device driver used to unconditionally reference
a libserialport header file. Remove that reference, it's not needed in
this specific driver.
Don't exit with an error if the FPGA is detected as unsupported.
Just issue a warning with the detected version and continue. I have such
a clone and it works with the original Saleae software and with sigrok
despite the fact that its FPGA version is 0xff.
Their commands are very similar to the U123x series, they just
add some more modes and the second channel. So use the re-functions
and just extend them where necessary.
Log reading not supported yet.
Basic testing done with a U1272A.
- fix the resolution of the CONF?-response:
the resolution is given with 6 decimal places
(instead of 8) like this:
VOLT +5.000000E+00,+1.000000E-04
- add more measurement modes that are possible with the meter:
CONT,COND,TEMP,PULS
src/hardware/sysclk-sla5032/protocol.c: In function ‘la_start_acquisition’:
src/hardware/sysclk-sla5032/protocol.c:244:8: warning: implicit declaration of function ‘max’ [-Wimplicit-function-declaration]
pre = max(pre, 2);
^~~
Without a measured quantity in packet.meaning->mq the C++ binding function
sigrok::Analog::mq() throws an exception and there is no way to check if
there is any measured quantity set in the analog package.
Add support for the Pepino-style of accessing >256K of memory. Because
this the only known extension of accessing >256K currently, we apply it
as soon as the sample size is bigger than 256K. Let's hope other
devices (if any) will follow this style. If not, we need to add support
depending on the device name later.
Signed-off-by: Wolfram Sang <wsa@the-dreams.de>
Magnus (creator of the Pipistrello) confirmed that he mixed up the
register names. The code was doing it correctly nonetheless but was
confusing to read because of this. Fix it to make it easier to
comprehend.
Signed-off-by: Wolfram Sang <wsa@the-dreams.de>
Let max_channels really carry the number of maximum channels the
hardware supports. We will handle the limitation of only half the
channels available in 200MHz mode later. Note that there won't be a
regression because we only set the variable but never check it. The
desired result of this patch is the removal of the NUM_CHANNELS macro.
The number of channels needs to be dealt with at runtime.
Signed-off-by: Wolfram Sang <wsa@the-dreams.de>
We needs this twice so put it into a seperate function, so updates to it
will automatically handled for both callers.
Signed-off-by: Wolfram Sang <wsa@the-dreams.de>
commit f51acd69 ("ols: combine demux samples") wrongly replaced the bit
pattern of 0x20 with the number of channels which just happens to be 32
as well. So, the code works but is confusing to read. Reword the
for-loop to make it more comprehensible.
Signed-off-by: Wolfram Sang <wsa@the-dreams.de>
The OLS protocol sends 16bit values to specify the sample count and
delay count. However, this 16bit value is the number of 32bit words to
be sampled, so the actual sample count is 4 times larger and does not
fit into a uint16_t. Extend it to support the full range of 256K
(LogicShrimp will need this) and to prepare support for devices with
even more memory (Pepino).
Signed-off-by: Wolfram Sang <wsa@the-dreams.de>
The previous implementation assumed that a receive data chunk ends
exactly with a sensor packet's end. Yet the buffer had 32 bytes while
the packets have 19 bytes.
Separate the data reception from the packet processing. Collect whatever
chunks the USB connection provides, and scan the resulting buffer for
packets. Cope with either incomplete or corrupt or misaligned packets as
well as with multiple packets in receive chunks. The latter might happen
upon initial synchronization, when a device already sends data or the
serial port buffered previously communicated data.
In the regular case, the computer will process so fast that each single
character will be handled individually. We don't mind. The frequency is
some 60 times per second, and the data volume is 19 bytes. The software
works for the regular case, and synchronizes fast at startup or after
comm errors.
Always print the data bytes of received buffers in the packet parser,
then check some more fixed fields to not process invalid packets, then
process the packet content as the previous implementation did.
Call the packet parser for incomplete packets and discarded input
buffers as well (initial synchronization, re-sync after comm errors).
This results in the availability of more diagnostics during development.
Pass the packet's location and size from outside. This prepares the
logic to cope with situations where the receive buffer contains multiple
(potentially incomplete) packets.
Slightly unobfuscate the UT32x packet parser. The protocol is mostly
ASCII based, checks for hex numbers may be unexpected. Use symbolic
identifiers for the packet length and some special characters.
The previous implementation of the UT32x driver expected to see a conn=
spec, without it no device is found. Default to the USB identification
of the CH9325 chip, to make the driver work out of the box. Users still
can provide conn= specs and override the default for other cables.
Fill in the scan, open/close, get/set/list, acquisition start/stop logic
such that data acquisition with a PICkit2 works.
Trigger support needs more attention. User specified triggers either
seem to not take effect, or the trigger position is not in the expected
location. It's yet to get determined what's the issue.
This implementation is based on protocol information gathered from the
pk2-la project.
Remove a free() call in an error path for a list which immediately
before the call was determined to be NULL. Use index 0 and 1 for
channels P1 and P2 respectively (the previous implementation used 0
for both channels).
The current implementation of the SCPI DMM driver is conservative about
checking the device's being operational, but the *OPC? queries are found
in unfortunate locations. Run the OPC query right before running the
next "actual" command, not afterwards. And certainly not between sending
requests and potentially gathering responses in subsequent calls.
This commit does not change current behaviour, but improves maintenance
before pending commits.
The "get MQ" helper routine communicates SCPI responses and translates
them to internal "MQ and flag" values. Optionally return the MQ table
entry reference to callers, so they don't have to repeat the table
lookup when the function's default precision is required, or should
future "start acquisition" requests need to refer to the meter's current
function.
Supported SCPI DMM devices will differ in the set of options and whether
parameters can get queried or configured. Use a "generic" set of devopts
during scan and for simpler models, prepare support for other sets of
devopts for more complex models.
Implement the scpi-dmm driver in such a generic way that it could work
with several protocol variants and with differing models which happen to
use any of these protocol variants. Prepare a list of supported models
with their respective SCPI command set, set of DMM functions and their
precision.
Add support for Agilent 34405A. The ten functions of this device got
tested and are operational, in continuous mode as well as with sample
count or capture time limits. The driver can query the current meter's
function, can change the function, and can run acquisitions in either
the current mode or with a user specified function selection. There is
some potential for improvement: AUTO/MIN/MAX/HOLD indicators are not
supported by this implementation.
Free the SCPI hardware info after successful model detection, too. Only
allocate the device instance when a supported model was found. Link the
device context earlier right after allocation, for easier verification.
The Fluke 45 probe routine tries to detect whether the serial port is
"in echo mode" (which already is questionable before the IDN query).
In the absence of a response, the library segfaults. Fix it.